MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Event Description
It was reported to philips that the protective cover on the light fell.The device was not in clinical use at the time of the event.No harm to the patient or user was reported.A philips field service engineer (fse) re-installed the cover which returned the device to use in good working order.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the arm support protection cover fell off.Upon troubleshooting, the fse reinstalled the protection cover.After that, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18510544
MDR Text Key333008119
Report Number3003768277-2024-00191
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838054189
UDI-Public00884838054189
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722026
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/27/2023
Initial Date FDA Received01/15/2024
Supplement Dates Manufacturer Received03/06/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse