Model Number ATEC SAPPHIRE 100
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2024
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that on (b)(6) during an atec procedure the tech heard a leak during the procedure and the cores were too small for diagnostic use.System was rebooted and the needle reinserted and the procedure could not be completed.System was examined and passed all tests, no suction issues found.Patient will need to return for 2nd procedure to obtain the samples.No additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATEC SAPPHIRE
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key18511140
MDR Text Key332899110
Report Number1222780-2024-00017
Device Sequence Number1
Product Code KNW
UDI-Device Identifier15420045504028
UDI-Public(01)15420045504028(11)170203
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/03/2017
Device Model NumberATEC SAPPHIRE 100
Device Catalogue NumberATEC SAPPHIRE 100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received01/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-