Model Number 77424
Device Problems Fluid/Blood Leak (1250); Misassembled During Installation (4049)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 12/16/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use of an eopa arterial cannula, it was reported that during cannulation of the ascending aorta for a cardiopulmonary bypass case, the customer described that the white internal obturator did not fit the device as usual.The customer described that the blood pressure in the aorta was pushing out the obturator and the device body filled with blood more rapidly than usual.There was no resistance in withdrawing the obturator compared to the customer's usual experience and there was some leakage of blood from the red cap/ white obturator junction. there was no damage to any packaging and the device did not look abnormal.The customer was a long-term user of the device.The device was used to complete the procedure, as it was already in the aorta and there were no patient issues and remainder of the operation continued without an event. there was no adverse patient effect associated with this event.Medtronic received additional information that the cannula was not heated or cooled prior to use.The patient blood loss was minimal (10 ml) but was scavenged by the heart lung machine and returned to the patient.A transfusion was not required as a result of this leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18516152
MDR Text Key332873025
Report Number2184009-2024-00017
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00763000135645
UDI-Public00763000135645
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K031037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number77424
Device Catalogue Number77424
Device Lot Number2023070731
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2024
Initial Date FDA Received01/16/2024
Date Device Manufactured07/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-