C.R. BARD, INC. (BASD) -3006260740 POWERPORT TI L/P 6 CF INT WOSP ATT SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8716001 |
Device Problems
No Display/Image (1183); Poor Quality Image (1408)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, images were provided for review.The investigation of the reported event is currently underway.H10: d4: (expiry date: 11/2024).H11: section a through f:the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that sometime post a port placement, the ct mark allegedly failed to display under x-ray imaging.There was no reported patient injury.
|
|
Event Description
|
It was reported that sometime post a port placement, the ct mark allegedly failed to display under x-ray imaging.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Three chest x-rays were provided for review.It depict the device placed via a right jugular access.The port has a triangular radiopaque housing, three open suture ports and does not have a ct marker.Therefore the investigation is confirmed for the reported poor quality image issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 11/2024), g3, h6 (method) h11: h6 (device, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|