• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIPATH IMAGING, INC VIAL SUREPATH COLLECTION KIT 500; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIPATH IMAGING, INC VIAL SUREPATH COLLECTION KIT 500; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED Back to Search Results
Catalog Number 491452
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
It was reported when using the vial surepath collection kit 500 the vial cap arrived cracked containing patient sample causing the sample to leak.No health impact or consequence reported.
 
Manufacturer Narrative
E1: initial reporter address: (b)(6).H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
G.4: date received by manufacturer has been corrected to 26dec2023.
 
Event Description
It was reported when using the vial surepath collection kit 500 the vial cap arrived cracked containing patient sample causing the sample to leak.No health impact or consequence reported.
 
Manufacturer Narrative
H.6.Investigation summary: the customer complaint is for one (1) vial with a broken cap (post sample collection) from item 491452 lot number 3060223.Material 491452 is produced at the bd mebane, nc facility on a validated automated manufacturing line.The capper section caps the vials to a validated application torque.The capper is validated to inspect for application torque and unseated or missing caps.Vials that fail to meet inspection requirements (i.E., outside of the validated application torque) are rejected automatically after the capper section.To ensure that the capper remains in validated state, a quarterly preventive maintenance (pm) is established that is used to confirm accuracy of application torque.The pm is performed by using a calibrated torque verifier that is compared against the application torque value.Production of material 491452 lot 3060223 started on 07mar2023.A review of the two (2) pm events that bracketed the production date identified that the results of the verification were acceptable.A total of 800 vials were qc inspected prior to product disposition with a total of 0 defects observed.An additional 800 vials were leak tested in a vacuum chamber during in-process testing and did not identify any leaking or cracked cap defects.The review of the manufacturing dhr for the lot number identified that it was complete and accurate with no indication of abnormal activities during manufacturing.The review of the bill of materials (bom) for 491452 lot 3060223 identified that raw cap lot number material 700030951 lot number 3017888 was used during the production.A review of the incoming inspection results for 700030951 lot number 3017888 identified 800 caps were inspected from the lot and passed the acceptance criteria with zero defects identified.A sample was not returned to the facility for evaluation.A visual retain analysis was performed on one clamshell (25 vials) from item 491452 lot 3060223.No cracked/broken caps were identified during the retain analysis.A picture was provided that shows a cracked cap.The complaint is confirmed.A 12-month complaint review for the defect mode of broken caps was performed and identified previous complaints for the item number and a previous broken cap complaint for the lot number.Bd performs regular trending to determine if a corrective and preventative action (capa) is required, and as of this time the threshold for a capa has not been reached.Bd will continue to monitor and evaluate trends.
 
Event Description
It was reported when using the vial surepath collection kit 500 the vial cap arrived cracked containing patient sample causing the sample to leak.No health impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIAL SUREPATH COLLECTION KIT 500
Type of Device
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer (Section G)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18519261
MDR Text Key332939156
Report Number3008007472-2024-00002
Device Sequence Number1
Product Code MKQ
UDI-Device Identifier10382904914520
UDI-Public(01)10382904914520
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
P970018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number491452
Device Lot Number3060223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received01/16/2024
Supplement Dates Manufacturer Received12/26/2023
02/20/2024
Supplement Dates FDA Received01/20/2024
03/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-