BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724E
Device Problem Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Event Description
It was reported that during use of two (2) em2400 valve sets, there was a large number of bubbles from ports 1 and 2.This was observed during filling of the bags.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter address: (b)(6).E1: initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correcton: d1, d4.Additional information: g4, h3, h4, h6, h11.H11: the actual devices were not available; however, six (6) companion samples were received for evaluation.Unaided eye visual inspection was performed on a returned sample and no defects were observed on the valve body cavity.System level and functional leak testing were performed on the sample using an in-lab compounder and no issue of bubbles were observed or encountered during testing throughout different compounding stages.The reported condition was not verified on the companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt 001 mz 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18519423
MDR Text Key333256328
Report Number1416980-2024-00050
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724E
Device Lot Number60372884
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2023
Initial Date FDA Received01/16/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-