Upon analysis during manufacturing, it was discovered that the cutting wire of the returned device had separated at the tip, with molten areas evident on both sides of the separation.Although the electrode tip showed melting on both sides of the separation without any abrasion, overheating is likely the cause.This overheating could have occurred due to various factors, such as accidental contact with another instrument causing a shortcut, prolonged operation without adequate cooling time, or selecting too high a voltage on the hf generator.It's important to note that the corresponding instruction for use (ifu 97000160_v 10.4) already includes a voltage limitation to prevent thermal overload.Furthermore, the ifu emphasizes that using device settings exceeding the specified values can lead to damage to the instrument.There have been no reports of patient harm or injury associated with this issue.The broken part was retrieved without any harm to the patient, therefore this case is non-reportable in united kingdom.Internal karl storz reference number: (b)(4).
|