ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY DR; Pulse generator, permanent, implantable Back to Search Results
Model Number PM2160
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 01/12/2024
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found, as received, the device output and telemetry communication were normal.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Visual inspection of the header attachment area detected a bonding anomaly.The device was tested for a hermeticity breach which was not observed.The device was cut open to enable further testing and battery was found normal.The hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18520485
MDR Text Key333133840
Report Number2017865-2024-01276
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507059
UDI-Public05414734507059
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2016
Device Model NumberPM2160
Device Lot Number4879472
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-