MEDTRONIC IRELAND IN.PACT PACIFIC; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER Back to Search Results
Catalog Number PCF05012013P
Device Problems Defective Device (2588); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  malfunction  
Event Description
It was reported with limited information that an in.Pact pacific catheter was defective.When the device was returned to the manufac turing facility, it was noted that there was a twist in the balloon.
 
Manufacturer Narrative
Product analysisno deformation was evident to the catheter body.The device returned with balloon folds expanded.Blood was visible in the balloon.A kink was evident to the distal section of the balloon.On pressurisation of the device, the balloon maintained pressure.Upon inspection, creases were visible on the balloon characteristic of a balloon twist.No deformation evident to the distal tip.No other damage evident to the remainder of the device.Please note that this device (in.Pact pacific) is not marketed in the united states; however, it is similar to the united states marketed product (in.Pact admiral).This event is being reported only as a malfunction because of the similar device requirement in 803 which is limited to malfunctions.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT PACIFIC
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18524048
MDR Text Key333003789
Report Number9612164-2024-00259
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P140010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPCF05012013P
Device Lot Number0011449867
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2024
Initial Date FDA Received01/17/2024
Supplement Dates Manufacturer Received01/17/2024
01/16/2024
Supplement Dates FDA Received01/22/2024
02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-