Model Number 16-02-85 |
Device Problems
Insufficient Cooling (1130); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in little rock, arkansas.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.It as possible to hear compressor relay turn on and off in both service mode, automatically by adjusting the set point and powering the unit on/off.It was suggested to replace the unit.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t did not cool both patient and cardioplegia sides during procedure; all other features worked properly.There was no patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: complaints database review revealed that no similar issue have been submitted for this unit since its installation in 2013.Based on the collected information and on investigation results of similar complaints, the root cause of the reported issue could be traced back to a degradation of cooling coil (coil micro-hole formation) leading to refrigerant leak and water entering the refrigeration cycle damaging the compressor.
|
|