• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/48; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/48; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 7236-2-848
Device Problems Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Patient Problem Joint Dislocation (2374)
Event Date 12/28/2023
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer. should additional information become available it will be reported in a supplemental report upon completion of the investigation.H3 other text : device not returned to the manufacturer.
 
Event Description
Rev ltha from dislocated mdm to constrained liner.
 
Manufacturer Narrative
Reported event: an event regarding disassociation involving an adm liner was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the ceramic head from the poly adm liner.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It is likely that patient trauma contributed to the device failure as it was reported that fall was involved.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Rev ltha from dislocated mdm to constrained liner.Update: femoral head disassociated from the poly liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORATION ADM X3 INS 28/48
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18524258
MDR Text Key332974369
Report Number0002249697-2024-00099
Device Sequence Number1
Product Code LZO
UDI-Device Identifier07613327315974
UDI-Public07613327315974
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182468
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number7236-2-848
Device Lot Number99715701
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/17/2024
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received03/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age85 YR
Patient SexFemale
Patient Weight75 KG
-
-