MAUDE Adverse Event Report: ABBOTT GMBH ARCHITECT ANTI-HCV REAGENT KIT; ASSAY, ENZYME LINKED IMMUNOSORBENT, HEPATITIS C VIRUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ARCHITECT ANTI-HCV REAGENT KIT; ASSAY, ENZYME LINKED IMMUNOSORBENT, HEPATITIS C VIRUS Back to Search Results
Catalog Number 06C37-74
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
A literature article by li h, yang s, cao d, wang q, zhang s, zhou y, liu d, yang r, cui l and zhu z (2023), ¿a new double-antigen sandwich test based on the light-initiated chemiluminescent assay for detecting anti-hepatitis c virus antibodies with high sensitivity and specificity¿, frontiers in cellular and infection microbiology 13:1222778, documented false nonreactive architect anti-hcv results generated on the architect i2000sr processing module for 6 patient samples.The samples were tested with other methods and the anti-hcv results were confirmed positive.The following data was provided in the article: architect anti-hcv result = 0.27 s/co (nonreactive) lica 500 system result = 3.07 s/co (reactive) roche cobas result = 14.33 s/co (reactive) riba 3.0 result = positive (bands on riba = core1, core2, ns3, ns4) hcv rna = negative anti-hcv confirmatory result = positive architect anti-hcv result = 0.17 s/co (nonreactive) lica 500 system result = 1.70 s/co (reactive) roche cobas result = 12.19 s/co (reactive) riba 3.0 result = positive (bands on riba = core1, core2, ns3, helicase) hcv rna = negative anti-hcv confirmatory result = positive architect anti-hcv result = 0.18 s/co (nonreactive) lica 500 system result = 1.38 s/co (reactive) roche cobas result = 11.23 s/co (reactive) riba 3.0 result = positive (bands on riba = core1, core2, ns3, ns5) hcv rna = negative anti-hcv confirmatory result = positive architect anti-hcv result = 0.12 s/co (nonreactive) lica 500 system result = 1.20 s/co (reactive) roche cobas result = 6.42 s/co (reactive) riba 3.0 result = positive (bands on riba = core1, core2, ns3) hcv rna = negative anti-hcv confirmatory result = positive architect anti-hcv result = 0.94 s/co (nonreactive) lica 500 system result = 54.73 s/co (reactive) roche cobas result = 59.66 s/co (reactive) riba 3.0 result = indeterminate (bands on riba = ns3, ns4) hcv rna = negative anti-hcv confirmatory result = positive architect anti-hcv result = 0.05 s/co (nonreactive) lica 500 system result = 12.11 s/co (reactive) roche cobas result = 26.17 s/co (reactive) riba 3.0 result = indeterminate (bands on riba = ns3, ns4) hcv rna = 857 iu/ml anti-hcv confirmatory result = positive there was no impact to patient management reported.
 
Manufacturer Narrative
This report is being filed on an international product, architect anti-hcv, list number 06c37-74, that has a similar product distributed in the us, list number 01l79.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
A literature article by li h, yang s, cao d, wang q, zhang s, zhou y, liu d, yang r, cui l and zhu z (2023), ¿a new double-antigen sandwich test based on the light-initiated chemiluminescent assay for detecting anti-hepatitis c virus antibodies with high sensitivity and specificity¿, frontiers in cellular and infection microbiology 13:1222778, documented false nonreactive architect anti-hcv results generated on the architect i2000sr processing module for 6 patient samples.The samples were tested with other methods and the anti-hcv results were confirmed positive.The following data was provided in the article: architect anti-hcv result = 0.27 s/co (nonreactive).Lica 500 system result = 3.07 s/co (reactive).Roche cobas result = 14.33 s/co (reactive).Riba 3.0 result = positive (bands on riba = core1, core2, ns3, ns4).Hcv rna = negative.Anti-hcv confirmatory result = positive.Architect anti-hcv result = 0.17 s/co (nonreactive).Lica 500 system result = 1.70 s/co (reactive).Roche cobas result = 12.19 s/co (reactive).Riba 3.0 result = positive (bands on riba = core1, core2, ns3, helicase).Hcv rna = negative.Anti-hcv confirmatory result = positive.Architect anti-hcv result = 0.18 s/co (nonreactive).Lica 500 system result = 1.38 s/co (reactive).Roche cobas result = 11.23 s/co (reactive).Riba 3.0 result = positive (bands on riba = core1, core2, ns3, ns5).Hcv rna = negative.Anti-hcv confirmatory result = positive.Architect anti-hcv result = 0.12 s/co (nonreactive).Lica 500 system result = 1.20 s/co (reactive).Roche cobas result = 6.42 s/co (reactive).Riba 3.0 result = positive (bands on riba = core1, core2, ns3).Hcv rna = negative.Anti-hcv confirmatory result = positive.Architect anti-hcv result = 0.94 s/co (nonreactive).Lica 500 system result = 54.73 s/co (reactive).Roche cobas result = 59.66 s/co (reactive).Riba 3.0 result = indeterminate (bands on riba = ns3, ns4).Hcv rna = negative.Anti-hcv confirmatory result = positive.Architect anti-hcv result = 0.05 s/co (nonreactive).Lica 500 system result = 12.11 s/co (reactive).Roche cobas result = 26.17 s/co (reactive).Riba 3.0 result = indeterminate (bands on riba = ns3, ns4).Hcv rna = 857 iu/ml.Anti-hcv confirmatory result = positive.There was no impact to patient management reported.
 
Manufacturer Narrative
The complaint investigation false nonreactive architect anti-hcv results included a review of data and information from the article ¿a new double-antigen sandwich test based on the light-initiated chemiluminescent assay for detecting anti-hepatitis c virus antibodies with high sensitivity and specificity¿, ticket trending review, device history record review, and labeling review.Return testing was not completed as returns were not available.The data and information provided in the article were reviewed and support the complaint issue.A ticket search by lot could not be performed as the likely cause lot is unknown.A review of the complaint trending report did not identify any trends for the issue for the product.A review of the device history record did not identify any non-conformances or deviations with the likely cause list number and the complaint issue.A review of labeling was performed and found to sufficiently address the customer's issue.Based on this investigation, no systemic issue or deficiency was identified with the architect anti-hcv reagent, lot number unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT ANTI-HCV REAGENT KIT
Type of Device
ASSAY, ENZYME LINKED IMMUNOSORBENT, HEPATITIS C VIRUS
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18525436
MDR Text Key333342065
Report Number3002809144-2024-00012
Device Sequence Number1
Product Code MZO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06C37-74
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/01/2024
Initial Date FDA Received01/17/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, (B)(6) ; ARC I2000SR INST, 03M74-02, (B)(6)
-
-