During the wolfe parkinson white procedure, noise was noted causing a delay.When the ablation catheter was inserted into the patient¿s right atrium, electrode 2 was noted to be off the ensite x screen and not visualized appropriately.Electrode 2 of the ampere connect cable was not displaying properly.There were no egms on the non-abbott (cardiolab) system.All connections from the catheter to the tactisys, and tactisys to the generator were reseated.The tactisys, tactiflex rf cable and catheter were replaced, which did not resolve the issue.The ampere connect cable was replaced to continue the procedure with no consequences to the patient.
|