MEDOS INTERNATIONAL SARL TRUESPAN 24 DEGREE PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 228152 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.(b)(4) e1: reporter facility: (b)(6) hospital.
|
|
Event Description
|
It was reported by the sales rep from china that during a meniscal repair surgical procedure on (b)(6) 2023 the truespan 24 degree peek device could not fire.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, a video was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned video.Upon visual inspection of the video, the video shows the device sitting on the packaging tray.The red trigger was found in its full retracted position and the pushing rod was found sticking out of the needle similar as the firing position; therefore, this complaint can be confirmed.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.A manufacturing record evaluation was performed for the finished device, and no non conformances were identified.Based on the investigation findings, the reported complaint was confirmed.Hands on analysis should provide the evidence necessary to confirm the root cause.The video does not provide enough evidence to determine root cause.Without physical evaluation of the device reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause for deployment issues could be related when not inserting the needle to the proper depth for deployment which have caused blocking insertion of the first implant, and when this occurred may have felt resistance and did not pull the trigger all the way.Also, the trigger might have not fully been pressed until a click sound was heard which could cause the reported failure.However, it cannot be conclusively affirmed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary the product was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned device.Visual inspection found that truespan 24 degree peek had the first implant was deployed.It does not show structural anomalies.The plates and the suture are in good conditions.The red trigger was found in a normal shape.No other anomalies could be observed.A functional test was performed to the second plate.The applier needle was introduced into a soft tissue simulator, the red trigger was fully squeezed, the second plate was successfully deployed, no obstruction or difficulties while deploying were found.It was verified that the pusher shaft tip is sliding out completely from the applier needle.The overall complaint was confirmed as the observed condition of the device contribute to the complained device issue.¿ based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|