This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section e1 (email address).The device was returned to zoll medical corporation for evaluation.The customer's report was not replicated or confirmed.A review of the device log from the reported event date found seven charge and shock events at 30j.Out of the seven times, six times the user discharged using the shock button into a defib pad short (manual readiness test).The remaining time, the user manually shocked 30j into a patient impedance using the shock button.The device was put through extensive testing including bench handling, impedance, defibrillation cycling, and environmental chamber testing without duplicating the report.An internal inspection of the device found no discrepancies.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|