Catalog Number 20039E
Device Problem Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd alaris smartsite extension set separated the following information was received by the initial reporter with the following verbatim: tubing getting disconnected from picc line twice in 24hrs.Product: al20039e.
 
Manufacturer Narrative
Dhr: no lot provided in initial mdr - provided by customer.No product or photo was returned by the customer.The customer complaint of connection issues could not be verified due to the product not being returned for failure investigation.A device history record review for material# 20039e and lot# 23095677 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 12oct2023.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event Description
Material #: 20039e, batch#: 23095677.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE EXTENSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18539513
MDR Text Key333498942
Report Number9616066-2024-00044
Device Sequence Number1
Product Code FPA
UDI-Device Identifier27613203013820
UDI-Public(01)27613203013820
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960280
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number20039E
Device Lot Number23095677
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/18/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-