AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Manufacturer Narrative
He device was returned and evaluated.During the evaluation, it was determined that deterioration of the angulation mechanism due to rust.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, that the ureteto-reno videoscope was leaking.The issue occurred during reprocessing of the device.The device was returned for evaluation.During the device evaluation, deterioration of the angulation mechanism by rust was found.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during the evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the likely cause of the reported event is due to the a-wire was stuck in the a-coil due to corrosion caused by fluid invasion on the insertion section.The suggested event is detectable at the inspection prior to use that is stated in operation manual 3.2 inspection of the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18539608
MDR Text Key333206685
Report Number9610595-2024-01168
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received01/18/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-