Model Number V60
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer on the v60 indicating that the device¿s display is dim and flickering.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.No medical intervention was provided to the patient, nor was a delay to therapy noted.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The rse advised the customer of the latest version of the service manual, and that they likely have a 1st generation hydis display installed.The rse provided the customer the parts to update to a 2nd generation nec display.The investigation is ongoing.
 
Manufacturer Narrative
Multiple attempts have been made to try to obtain further information about this case, but the customer never responded.The complaint record will be processed for closure.If additional information is received the complaint file will be reopened and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18540264
MDR Text Key333205504
Report Number2518422-2024-03436
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009851
UDI-Public00884838009851
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/18/2024
Supplement Dates Manufacturer Received04/12/2024
Supplement Dates FDA Received04/15/2024
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-