RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer on the v60 indicating that the device¿s display is dim and flickering.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.No medical intervention was provided to the patient, nor was a delay to therapy noted.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The rse advised the customer of the latest version of the service manual, and that they likely have a 1st generation hydis display installed.The rse provided the customer the parts to update to a 2nd generation nec display.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Multiple attempts have been made to try to obtain further information about this case, but the customer never responded.The complaint record will be processed for closure.If additional information is received the complaint file will be reopened and a supplemental report will be submitted.
|
|