• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Event Description
It was reported that the device's tip broke off when staff went to use the device.The device was replaced and the procedure was reported to have been completed successfully.There was no patient injury or medical intervention reported.There was no additional information provided regarding extended procedure time.These are commonly used devices that are readily available.
 
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the blade was broken off and was returned with the device in a specimen cup.The results of the visual inspection determined that the reported event was confirmed for fractured blade.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: - too much force or torque applied to instrument, or grasping/pulling - incidental and prolonged activation against solid surfaces, such as bone or plastic - attempting to bend, sharpen, or otherwise alter the shape of the blade the instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key18540889
MDR Text Key333221321
Report Number0001056128-2024-00019
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public07613327346251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number15342245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-