Catalog Number 8065750914
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Event Description
A nurse reported that no gas was out from the pipe during vitrectomy surgery.The surgery was completed on the same day after replacing the product with another one.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.An infusion manifold was returned.The returned sample was visually inspected, and no obvious defects were found.A calibrated console representing the current software version was used to test the sample.The sample was tested with a lab stock posterior cassette.The sample could prime and pass intra ocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.The infusion pressure was measured at multiple set points and met specifications.Toggling the infusion and the fluid/air exchange (f/ax) modes, fluid and air flowed from the cassette to the infusion line continuously without any bubble in various settings in all sub modes.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM, ACCESSORY, INFUSION TUBING SET
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18541477
MDR Text Key333217380
Report Number1644019-2024-00126
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657509140
UDI-Public00380657509140
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750914
Device Lot Number16316U
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/26/2023
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-