Catalog Number 121882750 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the ceramic insert shattered upon impaction.Event outcome/how was it managed? washout and alternative implanted.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information was received and stated as follows: a.Was there a surgical delay? if yes, what is the duration of the delay? 15 minutes.B.Was/were there any adverse consequence/s that affected the patient because of the reported event? extended surgical time.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Dhr review: product code 121882750 work order (b)(4) was manufactured on 06-oct-2022.(b)(4) parts were manufactured per specification and all raw materials met specification.There was no material non-conformance associated with this lot.There was no material reprocessing reports (mrr) associated with this lot.Conclusion: due to no similar failures found in the dhr review, the root cause of the complaint cannot be determined.Should further information come available that impacts the findings in this investigation it will be reopened.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : dhr review: product code 121882750 work order (b)(4) was manufactured on 06-oct-2022.(b)(4) parts were manufactured per specification and all raw materials met specification.There was no material non-conformance associated with this lot.There was no material reprocessing reports (mrr) associated with this lot.
|
|
Search Alerts/Recalls
|
|