• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Endocarditis (1834); Fever (1858); Heart Failure/Congestive Heart Failure (4446)
Event Date 11/17/2023
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The investigation determined the reported endocarditis appears to be related to patient conditions.The heart failure and fever are cascading effects of the endocarditis.The reported hospitalization and medication required were results of case specific circumstance as the patient returned to the hospital due to the patient effects and antibiotics were administered intravenously.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that on (b)(6) 2023, a mitraclip procedure was performed to treat a degenerative mitral regurgitation (mr) with grade of 4.Two clips were implanted, reducing mr to grade 2.On (b)(6) 2023, the patient was presented with heart failure, fever, and endocarditis on both implanted mitraclip.Vegetations were noted on echocardiogram.Antibiotics were administered intravenously.The patient is known with liver cirrhosis and experienced a septic arthritis of right knee which potentially could explain the onset of the endocarditis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18543692
MDR Text Key333253490
Report Number2135147-2024-00297
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/15/2023
Device Catalogue NumberCDS0702-XTW
Device Lot Number20816R1076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received01/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-