• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number 8360-10
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  Injury  
Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap inc.That a prestige atra grasper dbl-act 5mm (part # 8360-10) was used during a laparoscopic appendectomy procedure on january 8 2024.According to the complainant the grasper broke during a procedure.All pieces were successfully retrieved.An x-ray was performed and confirmed no fragments were left inside the patient.An unspecified delay to the procedure was reported.No patient complications were reported as a result of this event.The adverse event is filed under aic reference (b)(4).
 
Manufacturer Narrative
Updated information: h2, h6 codes updated.A picture of the device failure was sent.The picture included visible evidence of damage.The jaws of the device were shattered/broken.The picture confirmed that the jaws of the prestige grasper was broken/shattered.An approved project is in place to further address issues relating to the prestige grapser jaws breaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
3773 corporate parkway
center valley, PA 18034
4847197287
MDR Report Key18545474
MDR Text Key333271872
Report Number2916714-2024-00001
Device Sequence Number1
Product Code NWV
UDI-Device Identifier04046955083374
UDI-Public4046955083374
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number8360-10
Device Catalogue Number8360-10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received02/13/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-