RESPIRONICS, INC. DREAMSTATION2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number CAX521H12C
Device Problems Backflow (1064); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2022
Event Type  malfunction  
Event Description
The manufacturer received information alleging a dreamstation 2 advanced auto cpap would not turn on and no flow.There was no report of smoke, flames, voids, or exposed ac wires.There was no report of patient harm or injury.There was no report of medical intervention.The device was returned to the manufacturer for investigation.During the evaluation of the device, the service technician was able to confirm the customer's complaint of the device will not turn on and no flow.The service technician found of thermal damage to the pca and the lcd display.Evidence of water ingress was also verified.The technician was unable to download the error log data.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION2ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18546004
MDR Text Key333276736
Report Number2518422-2024-03578
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCAX521H12C
Device Catalogue NumberCAX521H12C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2023
Initial Date Manufacturer Received 04/06/2023
Initial Date FDA Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-