• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 600239264
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2024
Event Type  malfunction  
Event Description
During a supraventricular ventricular tachycardia (svt) procedure, electrical interference issues required multiple troubleshooting measures and resulted in a procedural delay.It was noted that there was electrical interference on the d-2 ablation catheter electrode.The ablation catheter was replaced, and there was still electrical interference.Then the catheter connector cable and the ampere rf generator were replaced but the issue persisted.The 1804 cable was then plugged into the cathlink, and the right electric potential was displayed.The procedure was completed without adverse consequences to the patient.
 
Event Description
This report includes a corrected pma/510(k) number.
 
Manufacturer Narrative
Corrected data: b5, g3 additional information: d9, g3, h2, h3, h6 the results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18547719
MDR Text Key333392862
Report Number2184149-2024-00019
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K141050, K20
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number600239264
Device Lot Number8819478
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/07/2024
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received02/23/2024
Supplement Dates FDA Received02/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2023
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-