ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 600239264 |
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2024 |
Event Type
malfunction
|
Event Description
|
During a supraventricular ventricular tachycardia (svt) procedure, electrical interference issues required multiple troubleshooting measures and resulted in a procedural delay.It was noted that there was electrical interference on the d-2 ablation catheter electrode.The ablation catheter was replaced, and there was still electrical interference.Then the catheter connector cable and the ampere rf generator were replaced but the issue persisted.The 1804 cable was then plugged into the cathlink, and the right electric potential was displayed.The procedure was completed without adverse consequences to the patient.
|
|
Event Description
|
This report includes a corrected pma/510(k) number.
|
|
Manufacturer Narrative
|
Corrected data: b5, g3 additional information: d9, g3, h2, h3, h6 the results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
|
Search Alerts/Recalls
|
|
|