Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The tip-up fenestrated grasper (tufg) instrument was analyzed and found to have the main tube broken.A piece measuring proximately 0.201¿ x 0.145¿ was broken off the instrument.The instrument housing was removed, and the clamping pulley screws were loosened to separate the main tube from the proximal clevis.Visual inspection found all internal cables to be still intact but there is a possibility of missing material.There was a chunk of material missing, and the material was unable to be found on the inside of the main tube.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.Blank mdr fields: the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Fields g5 and g7 are not applicable.
|