C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1716070J |
Device Problem
Material Protrusion/Extrusion (2979)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 09/2024) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that sometime post a port placement, the septum was allegedly damaged and was floating upon removal.Reportedly, the port was removed.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport slim implantable port attached to a catheter was returned for evaluation.Visual, microscopic, tactile and functional evaluations were performed on the returned device.The port septum was noted to be partially dislodged.Therefore the investigation is confirmed for the reported material protrusion issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 09/2024), g3 h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|
Event Description
|
It was reported that sometime post a port placement, the septum was allegedly damaged with slightly curled up and was found to be floating upon removal.Reportedly, the port was removed.There was no reported patient injury.
|
|