Model Number 20014
Device Problem Mechanical Problem (1384)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 12/27/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: visual inspection showed that the introducer did not come apart at the luer connection.The reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use of a dlp aortic root cannula with vent line, the customer reported that they cannulated the aorta and they were attempting to remove the metal cannula portion with the needle but they were unable to twist the metal needle from the cannula leur lock.The customer was unsuccessful in trying to twist the needle even with a kelly instrument.Due to this event, the customer removed the entire cannula and had to re-cannulate with a new dlp device.There was an unknown amount of blood loss due to this event during the procedure.There was no damage noted to the packaging or shipping box of these cannulas.There was no long term affect to patient noted.The device was replaced to complete the procedure.There was no additional adverse patient effect associated with this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic received additional information that when asked was there a blood transfusion required as a result of the blood loss, the customer stated that upon review of the chart for this case the patient did have to receive 1 unit of packed red blood cells (prbcs).The initial cannulation site was used when the cannula was replaced.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DLP AORTIC ROOT ANTEGRADE ANTEGRADE CANNULA WITH VENT LINE
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18551573
MDR Text Key333342538
Report Number2184009-2024-00026
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00613994495549
UDI-Public00613994495549
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K810548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20014
Device Catalogue Number20014
Device Lot Number2023090252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/27/2023
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received01/22/2024
04/04/2024
Supplement Dates FDA Received02/14/2024
04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight137 KG
-
-