MAUDE Adverse Event Report: GE HEALTHCARE T-PIECE NEONATAL PATIENT CIRCUIT KIT; ATTACHMENT, BREATHING, POSITIVE END EXPIRATORY PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE HEALTHCARE T-PIECE NEONATAL PATIENT CIRCUIT KIT; ATTACHMENT, BREATHING, POSITIVE END EXPIRATORY PRESSURE Back to Search Results
Catalog Number M1091335VS
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2023
Event Type  malfunction  
Event Description
T-piece neonatal patient circuit kit, disposable, lot# 23207 manufacture date [redacted date] peep valve faulty.3 circuits used on 2 different panda resuscitators would only provide 3 cmh2o of peep despite turning valve higher and higher.Tested off patient and found that when turned as high as the peep will go peep jumped from 3-10 cmh2o.Saved circuits and package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PIECE NEONATAL PATIENT CIRCUIT KIT
Type of Device
ATTACHMENT, BREATHING, POSITIVE END EXPIRATORY PRESSURE
Manufacturer (Section D)
GE HEALTHCARE
3000 n. grandview blvd
waukesha WI 53188
MDR Report Key18551664
MDR Text Key333343973
Report Number18551664
Device Sequence Number1
Product Code BYE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberM1091335VS
Device Lot Number23207
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/01/2023
Event Location Hospital
Date Report to Manufacturer01/22/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/22/2024
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-