• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD (SUZHOU) BD PEGASUS YEL 24GA X 0.75IN PRN-CAP Y; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD (SUZHOU) BD PEGASUS YEL 24GA X 0.75IN PRN-CAP Y; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383912
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd pegasus yel 24ga x 0.75in prn-cap y adapter / connector defective / damaged the following information was provided by the initial reporter: when giving intravenous fluids to a patient, the connection between the indwelling needle and the infusion connector could not be tightened, after replacing it, it was normal, replaced with a new consumable, and it was used normally without causing any harm to the patient.
 
Event Description
No additional information provided.
 
Manufacturer Narrative
1.Dhr/bhr review: (1)the batch number of the complained product is 2347346, is 24g and product code is 383912, produced on 2023/01, with a total of (b)(4) pieces in this batch; (2)inspection process and delivery test report, test results meet product standards, no abnormalities; (3)check the production record of the batch of products, no conformance, deviation or rework activities in the process of the batch of products; 2.The customer did not return samples and photos, and the defect status cannot be confirmed.3.Take the retained sample of this batch for 45psi system leakage test, and no abnormality was found.Test report refer to attachment 1.4.The history of customer complaints for the same batch of products has been reviewed, and no complaints of the same defects have been found.In summary,due to the lack of samples returned by the customer, the specific defect status cannot be confirmed, therefore the root cause of the complaint cannot be confirmed.The factory will continue to monitor the trend of the defect complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PEGASUS YEL 24GA X 0.75IN PRN-CAP Y
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18553226
MDR Text Key333517126
Report Number3014704491-2024-00019
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383912
Device Lot Number2347346
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/02/2024
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-