It was reported that bd pegasus yel 24ga x 0.75in prn-cap y adapter / connector defective / damaged the following information was provided by the initial reporter: when giving intravenous fluids to a patient, the connection between the indwelling needle and the infusion connector could not be tightened, after replacing it, it was normal, replaced with a new consumable, and it was used normally without causing any harm to the patient.
|
1.Dhr/bhr review: (1)the batch number of the complained product is 2347346, is 24g and product code is 383912, produced on 2023/01, with a total of (b)(4) pieces in this batch; (2)inspection process and delivery test report, test results meet product standards, no abnormalities; (3)check the production record of the batch of products, no conformance, deviation or rework activities in the process of the batch of products; 2.The customer did not return samples and photos, and the defect status cannot be confirmed.3.Take the retained sample of this batch for 45psi system leakage test, and no abnormality was found.Test report refer to attachment 1.4.The history of customer complaints for the same batch of products has been reviewed, and no complaints of the same defects have been found.In summary,due to the lack of samples returned by the customer, the specific defect status cannot be confirmed, therefore the root cause of the complaint cannot be confirmed.The factory will continue to monitor the trend of the defect complaint.
|