Model Number V60 V60PLUS VENTILATOR
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating there was check vent: primary alarm failed error code displayed.The device was reported to be in use at the time of the reported problem.No patient or user harm reported.The remote service engineer (rse) informed the customer that when the primary audible alarm test fails for either speaker, the alarms are produced using both the primary and backup audio devices.The rse advised the customer to perform the following troubleshooting steps in this order: listen for audible sound from the speakers, verify that the speakers are connected, verify that the braided wires are not touching the speaker housing, verify the resistance of each speaker is within specification, replace speaker #1, replace speaker #2, replace the central processing unit (cpu) printed circuit board assembly (pcba).The rse advised the customer to replace the #1 speaker connected to the motor controller (mc) pcba and provided the speaker assembly part information.This investigation is ongoing.
 
Manufacturer Narrative
H10: multiple good faith efforts (gfe) were attempted to obtain confirmation of the part order, part replacement, and resolution.No response or further information was received from the customer.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18554776
MDR Text Key333389509
Report Number2518422-2024-03790
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received02/09/2024
Supplement Dates FDA Received02/09/2024
Date Device Manufactured02/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-