RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60 indicating that a field service engineer (fse) was on-site to implement fco66.After extracting the log, an error 110b proximal pressure sensor failed message was identified.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The rse dispatched an fse for onsite repair.This investigation is ongoing.
|
|
Manufacturer Narrative
|
After further troubleshooting, in accordance with the service manual, the field service engineer (fse) replicated the symptoms by conducting the pre-operational test.Although no errors were visible on the unit screen, an error with the code associated with proximal pressure sensor auto zero failed message was detected in the log.The fse replaced the data acquisition pcba board to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|