TAEWOONG MEDICAL CO.,LTD. NITI-S ESOPHAGEAL COVERED STENT; ESOPHAGEAL STENT Back to Search Results
Model Number EST2215F-22
Device Problems Activation, Positioning or Separation Problem (2906); Activation Failure (3270); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the delivery system was broken during deployment.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.However, it is hard to exactly analyze because the device was not returned yet.Investigation will be conducted once device is returned and we will send follow-up report accordingly.
 
Event Description
Stent would not deploy.Stent would not come out of sheath when deployment system being pressed.Deployment system then broke.A new stent needed to be gotten and used without difficulty.
 
Manufacturer Narrative
It was reported that stent would not deploy and deployment system then broke.As a result of analysis of returned device, the outer sheath was detached and stent was partially deployed.There was curve on the stent loaded part of the outer sheath, and deployment was tried without pressure, and very strong resistance occurred, but it was gradually deployed, resulting in deployment success.In the inner sheath, notable kinking was not observed.Resistance can be felt due to pressure generated by patient's lesion during deployment.It can cause deployment failure if deployment is tried by force in this situation since outer sheath can be stretched and detached.However, it is hard to identify the exact root cause since it is hard to reconstruct the situation at the time of procedure.Based on the deployment success under no pressure even though the strong resistance has occurred, detached outer sheath and curve on the stent loaded part, it is considered that delivery system was pressured due to patient's lesion during the procedure and deployment was tried in that situation.It was hard to deploy due to pressure/curved sheath, in which concentrated the force causing strong resistance and the outer sheath was detached in the end, resulting in deployment failure.In addition, it is supposed that strong resistance occurred because delivery system was deformed during procedure.This complaint is assumed that it was a malfunction of the device due to the pressure of the patient's lesion, there will be continued monitoring of the same or similar customer complaints.
 
Event Description
Stent would not deploy.Stent would not come out of sheath when deployment system being pressed.Deployment system then broke.A new stent needed to be gotten and used without difficulty.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ESOPHAGEAL COVERED STENT
Type of Device
ESOPHAGEAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key18560285
MDR Text Key333472567
Report Number3003902943-2024-00003
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221482
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEST2215F-22
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 01/10/2024
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received01/10/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-