MAQUET SAS POWERLED; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD568350933
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2023
Event Type  malfunction  
Event Description
On 2nd june, 2023 getinge became aware of an issue with one of surgical lights - powerled 500.As it was stated, the covers are missing.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination in case of reoccurrence.
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.E1h event site postal code: (b)(6).H3 other text : device not returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLED
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key18562074
MDR Text Key333447890
Report Number9710055-2024-00069
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodePO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568350933
Device Catalogue NumberARD568350933
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/02/2023
Initial Date FDA Received01/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/20/2013
Is the Device Single Use? No
Type of Device Usage Initial