Model Number 799-EMBP-05-01
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/25/2023
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "internal comm failure - 1173", "internal comm failure - 1471", and "internal comm failure - 1475" messages.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
The customer's report was observed during review of the forensic memory log.However, the device was put through extensive testing without duplicating the report.An internal inspection identified that a cpu board pin was found to be misaligned and as a result the cpu to pim board ribbon cable was not fully inserted which may have contributed to the customer report.The cpu pin was reworked to enable proper connection with the cpu to pim ribbon cable as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18565259
MDR Text Key333500636
Report Number1220908-2024-00108
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number799-EMBP-05-01
Device Catalogue Number799-EMBP-05-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received01/23/2024
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-