Model Number M3535A
Device Problems Crack (1135); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Event Description
It was reported to philips that the device had fallen to the ground and that the power module had a broken case as a result.There was no patient involvement.
 
Manufacturer Narrative
Philips received a complaint on the heartstart mrx indicating that the defibrillator has slight dents.The device was not in clinical use at the time the issue was discovered.The complaint investigation revealed that the defibrillator fell to the ground, resulting in slight dents, but the device still passed the functional test.Based on the available information and the conducted testing, the cause of the reported problem was likely that the defibrillator falling to the ground and sustaining damage occurred due to its lack of mechanical attachment to the cart it was placed on, possibly as a result of the cart's fast movement.The defibrillator still passed the functional test despite the slight dents.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key18565654
MDR Text Key333506031
Report Number3030677-2024-00319
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/10/2024
Initial Date FDA Received01/23/2024
Supplement Dates Manufacturer Received02/21/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-