Model Number 16-02-85 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: patient pump 2 did not turn when activated.In order to solve the issue, the entire patient pump bridge was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message associated to patient 2 pump blocked or too slow.The issue occurred during procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
H10: based on technical intervention and considering investigation results of previous similar events, the most likely root cause of the reported issue can be associated to pump motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.
|
|
Event Description
|
See initial report.
|
|