• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number FX2100X15B
Device Problems Communication or Transmission Problem (2896); Mechanical Jam (2983); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2023
Event Type  malfunction  
Event Description
A trilogy evo ventilator was returned to the manufacturer for routine preventative maintenance.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, critical error code e-81, e-84 and e-150 was found in the ventilator's downloaded error log.The device's oxygen blending module board and filter was replaced to address the issue.The device was returned the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18570752
MDR Text Key333602986
Report Number2518422-2024-04069
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959061019
UDI-Public00606959061019
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K181166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFX2100X15B
Device Catalogue NumberFX2100X15B
Initial Date Manufacturer Received 08/22/2023
Initial Date FDA Received01/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No