This report is being supplemented to provide additional information based on the legal manufacturer's investigation.D9, h3, and h6 have also been updated.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the physical damage observed on the device resulted from improper handling.However, the root cause of the events could not be determined.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: the device ifu (instruction for use_99-1088_ff) states: "study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects." (page 4); "pay close attention to the care, cleaning, disinfection, and sterilization instructions in this manual.Any deviation can cause damage." (page5).Olympus will continue to monitor the field performance of this device.
|