• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VITALIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VITALIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number K287
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  Injury  
Event Description
It was reported that this pacemaker reverted to safety mode, which, resulted in an alert in the remote home monitoring system.Technical services (ts) discussed device function in safety mode and recommended further review.No further assistance was requested at this time.No adverse patient effects were reported.This device remains in service.If pertinent information is provided in the future, a supplemental report will be submitted.It was reported that surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device determined it had undergone resets and that bradycardia therapy remained available.The system resets occurred during other transient elevated power states and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
 
Event Description
It was reported that this pacemaker reverted to safety mode, which, resulted in an alert in the remote home monitoring system.Technical services (ts) discussed device function in safety mode and recommended further review.No further assistance was requested at this time.No adverse patient effects were reported.This device remains in service.If pertinent information is provided in the future, a supplemental report will be submitted.It was reported that surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.The product has been received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALIO
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18572707
MDR Text Key333619368
Report Number2124215-2024-03602
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/01/2016
Device Model NumberK287
Device Catalogue NumberK287
Device Lot Number246500
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received01/24/2024
Supplement Dates Manufacturer Received04/30/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age94 YR
-
-