The investigation was performed based on the provided log files and the onsite examination of the affected fabius gs by the dräger service.Analysis of the log files showed that the device has shut-down automatic ventilation due to measured speed fluctuations at the ventilator.The vacuum pump and motor have been replaced by the dräger technician and were sent in for examination.The vacuum pump was found to work as specified but the motor was stuck at one position which was identified as root cause for the reported stop of ventilation.The affected motor is approximately 19 years old.Wear and tear of the collector disc may have caused the speed fluctuations during motor rotation.These speed fluctuations will result in deviations between the intended piston hub and the real one and, the applied tidal volumes will not match the settings.To prevent from potentially hazardous output the device is designed to shut down automatic ventilation and to post a corresponding alarm to alert the user.Manual ventilation and the monitoring functions remain available to the full extend which enables the user to finish the particular surgical procedure.After replacement of the ventilator motor and vacuum pump by dräger service the device was confirmed to be operating per manufacturer¿s specification.Dräger finally concludes that the device behaved as specified upon the malfunction of a single component after almost 19 years of use; no patient consequences have been reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|