• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. SUPRAMID BLACK 4/0 (1.5) 45CM DS12; OTHER SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. SUPRAMID BLACK 4/0 (1.5) 45CM DS12; OTHER SUTURES Back to Search Results
Model Number G0712086
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported an issue with the product supramid: the reporter indicated that the thread has come loose from the needle.No further information received.
 
Manufacturer Narrative
G4: reported device marketed in the u.S.Only for veterinary use, however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.Summary of investigation: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample for analysis.Without any closed sample we cannot carry out an analysis in order to take a decision.Batch manufacturing record: reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfils usp/ep and b.Braun surgical requirements.Conclusion root cause analysis: it has not been possible to determine the root cause because samples have not been received.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Corrective measures: actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPRAMID BLACK 4/0 (1.5) 45CM DS12
Type of Device
OTHER SUTURES
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa, 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa, 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa, 121
rubí, barcelona 08191
SP   08191
MDR Report Key18578682
MDR Text Key333700941
Report Number3003639970-2024-00023
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG0712086
Device Catalogue NumberG0712086
Device Lot Number622272
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/05/2024
Initial Date FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-