MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITAL SIGNS MONITOR Back to Search Results
Model Number EARLYVUE VS30 VITAL SIGNS MONITOR
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Manufacturer Narrative
A philips remote service engineer (rse) spoke to the customer and determined that the device came out of its support base when the nurses equipped the device with consumables.The customer reported weakness in the device's closing system on its mounting base.The rse recommended replacement of the front facade.Based on the information available and the testing conducted, the cause of the reported problem was the mount.The reported problem was confirmed.The customer will be provided a replacement front assembly to resolve the issue.It has been concluded that no further action is required at this time.(b)(6).
 
Event Description
Philips received a complaint on the earlyvue vs30 vital signs monitor indicating the screen was broken when the device fell.The device was in clinical use at the time the issue was occurred.There was no adverse event or patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EARLYVUE VS30 VITAL SIGNS MONITOR
Type of Device
EARLYVUE VS30 VITAL SIGNS MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18578879
MDR Text Key333722950
Report Number1218950-2024-00054
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838091412
UDI-Public00884838091412
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K190624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEARLYVUE VS30 VITAL SIGNS MONITOR
Device Catalogue Number863359
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2022