HOLOGIC, INC FLUENT FLUID MANAGEMENT SYSTEM; INSUFFLATOR, HYSTEROSCOPIC Back to Search Results
Model Number RM-FLT-100
Device Problem Use of Device Problem (1670)
Patient Problem Hypervolemia (2664)
Event Date 01/04/2024
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that during a fluent procedure on (b)(6) 2024, the physician resected a fibroid and exceeded the deficit limit of 2500 ml 3 times.It was reported that the deficit was met 3 times and, each time, a manual count was performed.It was reported that the patient had to be taken to the intensive care unit.The final manual count was reported at 5000 ml.No other information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENT FLUID MANAGEMENT SYSTEM
Type of Device
INSUFFLATOR, HYSTEROSCOPIC
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key18579384
MDR Text Key333727321
Report Number1222780-2024-00025
Device Sequence Number1
Product Code HFG
UDI-Device Identifier15420045508552
UDI-Public(01)15420045508552(11)230104
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberRM-FLT-100
Device Catalogue NumberRM-FLT-100
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient SexFemale
-
-