DEPUY MITEK LLC US GRYPHON P BR ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 210814 |
Device Problems
Break (1069); Device Damaged Prior to Use (2284)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Udi: (b)(4).Investigation summary : a video was returned to depuy synthese mitek for evaluation.The depuy synthese mitek team conducted a visual inspection of the returned video.Upon visual inspection of the video, the device was observed on the package tray, the anchor was detached from the inserter, the inserter does not appear to have any structural anomaly.The suture and the anchor appear to have foreign matter, presumably biological matter.The anchor condition cannot be clearly observed; therefore, this complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device lot number: 160l908, and no nonconformances were identified.As part of depuy synthese mitek quality process all devices are manufactured, inspected, and released to approved specifications.Based on the investigation findings, the reported complaint was not confirmed.The video does not provide enough evidence to determine root cause.Without physical evaluation of the device reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause for the foreign matter on the anchor and suture could be related to the use of the device during the procedure in contrast with the event reported.However, it cannot be conclusively affirmed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthese mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
Event Description
|
It was reported by the sales rep in china that during an unknown procedure on (b)(6) 2024, it was observed that the anchor on the gryphon p br anchor w/oc device was broken off upon opening its package.During in-house engineering evaluation of the video provided by the customer, it was determined that the suture and the anchor on the device appeared to have foreign matter, presumably biological matter.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Investigation summary visual observations revealed that the device does not show structural anomalies, the suture and the anchor were returned for evaluation, the anchor was broken near the distal end.Also, foreign material was noted in the suture and anchor, presumably biological matter.The overall complaint was confirmed as the observed condition of the gryphon p br anchor w/oc would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedure variables, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|