E1: establishment name: (b)(6) hospital, independent administrative agency, regional medical function promotion.The device was returned to olympus for evaluation and the evaluation found the device passed all functional testing.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, as well as corrections to b3.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the noise or b30 mentioned on the actual product were not reproduced.It is possible that the cause may be a temporary problem such as an abnormality in the power supply environment, peripheral equipment (system, etc.), or poor contact with the scope.However, since the event was not reproduced, it was difficult to conduct further analysis on the actual product, and the cause was unable to be identified.Olympus will continue to monitor field performance for this device.
|