Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator, indicating that there was a check vent: proximal pressure sensor auto zero failed alarm.The device was reported to be outside of use at the time of the reported problem.No patient or user harm reported.The customer informed the remote service engineer (rse) of the error code and then confirmed the alarm in the device event log.The customer stated that no work had been performed on the device.The rse advised the customer to perform the following troubleshooting steps in this order: verify the pin alignment between the solenoids and data acquisition (da) printed circuit board assembly (pcba), replace the 3rd and 4th proximal autozero solenoids, and then replace the da pcba.The customer stated that they performed a full performance verification test on the device and was unable to duplicate the reported problem.The rse advised the customer if the same issue ever returned to follow the troubleshooting steps previously provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18580154
MDR Text Key333733383
Report Number2518422-2024-04306
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received01/25/2024
Date Device Manufactured11/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-