• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Event Description
It was reported that during device unpacking, the actuator knob was discovered to be detached from the system.There was no patient involvement.The device was replaced and the procedure was completed.
 
Manufacturer Narrative
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Manufacturer Narrative
All available information was investigated, and the reported detached actuator knob was confirmed via returned device analysis.Additionally, the release crimper was observed to be scratched.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the cause of the observed scratched release crimper was unable to be determined.The investigation determined the reported detached actuator knob to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18580958
MDR Text Key333739782
Report Number2135147-2024-00355
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/20/2024
Device Catalogue NumberCDS0702-XTW
Device Lot Number30522R1117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/25/2024
Supplement Dates Manufacturer Received03/18/2024
Supplement Dates FDA Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-