ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B12SRT
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4) date sent: 1/25/2024 d4: batch # unk this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown surgery, after inserting into the trocar, noted the device was broken.Another device was used to complete the surgery.There was no patient consequence reported.No additional information could be provided.
 
Manufacturer Narrative
(b)(4).Date sent: 2/13/2024.D4: batch # a9d109.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that b12srt device was received with the obturator inserted thru the sleeve.Upon evaluation of the device, the clear lens from the obturator were found damage.The sleeve was found without any damaged.The damaged on the clear lens, it is possible that the damaged was due to an improper handling of the device.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18581190
MDR Text Key333931375
Report Number3005075853-2024-00721
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001188
UDI-Public10705036001188
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB12SRT
Device Lot Number465C12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/27/2023
Initial Date FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-