A steris surgical service representative inspected the lighting system and reported the indicated yoke cover was damaged as the mounting screws and parts of the yoke cover were still attached to the lighting system.The observed damage is indicative of impact damage caused by user facility personnel bumping the light head into walls or other equipment.The harmony led surgical lighting operator manual is included with every steris harmony led surgical lighting system sold.The following cautionary statement is documented within the safety precautions section of this manual.The instructions further state (1-3), "caution - possible equipment damage hazard: do not bump light heads into walls or other equipment.Always use handles when positioning light head during surgical procedures, or when cleaning or servicing the lighting system." a steris account manager contacted the customer to inquire an in-service training presentation on the use of their harmony led surgical lighting system, specifically not bumping the lighting system into walls or other equipment, however the customer declined.No additional issues have been reported.
|
The user facility reported that during a patient procedure, the plastic lighthead yoke cover on the harmony led surgical lighting system detached and fell from above, causing the cover to contact a surgeon on the head and shoulder.No report of injury.The procedure was completed successfully.
|